EasyManuals Logo

Rigol MSO1074Z Programming Guide

Rigol MSO1074Z
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #169 background imageLoading...
Page #169 background image
Chapter 2 Command System RIGOL
MSO1000Z/DS1000Z Programming Guide 2-153
:TIMebase Commands
The :TIMebase commands are used to set the horizontal parameters, such as enabling the delayed sweep
and setting the horizontal timebase mode.
Command List:
:TIMebase:DELay:ENABle
:TIMebase:DELay:OFFSet
:TIMebase:DELay:SCALe
:TIMebase[:MAIN]:OFFSet
:TIMebase[:MAIN]:SCALe
:TIMebase:MODE
:TIMebase:DELay:ENABle
Syntax
:TIMebase:DELay:ENABle <bool>
:TIMebase:DELay:ENABle?
Description
Enable or disable the delayed sweep, or query the status of the delayed sweep.
Parameter
Name
Type
Range
Default
<bool>
Bool
{{1|ON}|{0|OFF}}
0|OFF
Explanation
Delayed sweep can be used to enlarge a length of waveform horizontally to view
waveform details.
Return Format
The query returns 1 or 0.
Example
:TIMebase:DELay:ENABle ON /*Enable the delayed sweep*/
:TIMebase:DELay:ENABle? /*The query returns 1*/

Table of Contents