RIGOL Chapter 2 Command System
2-168 MSO1000Z/DS1000Z Programming Guide
:TRIGger:SLOPe
Command List:
:TRIGger:SLOPe:SOURce
:TRIGger:SLOPe:WHEN
:TRIGger:SLOPe:TIME
:TRIGger:SLOPe:TUPPer
:TRIGger:SLOPe:TLOWer
:TRIGger:SLOPe:WINDow
:TRIGger:SLOPe:ALEVel
:TRIGger:SLOPe:BLEVel
:TRIGger:SLOPe:SOURce
Syntax
:TRIGger:SLOPe:SOURce <source>
:TRIGger:SLOPe:SOURce?
Description
Set or query the trigger source in slope trigger.
Parameter
Name
Type
Range
Default
<source>
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
CHANnel1
Return
Format
The query returns CHAN1, CHAN2, CHAN3, or CHAN4.
Example
:TRIGger:SLOPe:SOURce CHANnel2 /*Set the trigger source to CH2*/
:TRIGger:SLOPe:SOURce? /*The query returns CHAN2*/