Chapter 2 Performance Verification Test RIGOL
RSA5000 Performance Verification Guide
2-7
Absolute Amplitude Accuracy Test
Specification
Absolute Amplitude Accuracy
f
c
= 50 MHz, peak detector, PA off, attenuation = 10 dB, input signal level = -10 dBm, 20℃ to 30℃
Uncertainty
<0.3 dB
Test Devices
1. Signal Generator × 1
2. Power Meter × 1
3. Power Sensor × 1
4. Power Meter Connecting Cable × 1
5. Dual-N Cable × 1
6. Dual-BNC Cable × 1
Test Connection Diagram
(a)
(b)
RSA5000
Dual-N Cable
Dual-BNC
Cable
RF IN
RF OUT
Dual-N Cable
RF OUT
Power Meter
Power Sensor
Power Meter
Connecting
Cable
10MHz OUT
10MHz IN
Signal Generator
Signal Generator
Figure 2-3 Absolute Amplitude Accuracy Test Connection Diagram