Chapter 2 Performance Verification Test RIGOL
RSA5000 Performance Verification Guide
2-19
1 dB Gain Compression Test
Specification
1 dB Gain Compression
[1]
fc ≥ 50 MHz, attenuation = 0 dB, preamp off
Specification
0 dBm (nominal)
Note:
[1]
The frequency interval of the two-tone signals should be greater than 10 MHz.
Test Devices
1. Signal Generator × 2
2. Power Meter × 1
3. Power Sensor × 1
4. Power Meter Connecting Cable × 1
5. Power Divider × 1
6. Dual-N Cable × 3
7. Dual-BNC Cable × 1