EasyManua.ls Logo

Rigol RSA5000 Series - Page 24

Rigol RSA5000 Series
43 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL Chapter 2 Performance Verification Test
2-12 RSA5000 Performance Verification Guide
PA On:
Reference
Value 1
Reference
Value 2
Signal
Generator
Output
Frequency
Power Meter
Measurement
Value A1
Spectrum
Analyzer
Measurement
Value A2
System
Error
[1]
Global
Error
[2]
Frequency
Response
[3]
Limit
Pass/Fail
100 kHz
100
kHz to
3.2
GHz:
<0.7 dB,
<0.3 dB
(typical)
3.2 GHz
to 6.5
GHz:
<0.9 dB,
<0.5 dB
(typical)
500 MHz
2 GHz
3 GHz
4 GHz
5 GHz
6 GHz
6.499995
GHz
Note
[1]
: System Error = Power Meter Measurement Value A1 - Reference Value 1
Note
[2]
: Global Error = Spectrum Analyzer Measurement Value A2 - Reference Value 2
Note
[3]
: Frequency Response = |Global Error - System Error|

Other manuals for Rigol RSA5000 Series

Related product manuals