Appendix RIGOL
RSA5000 Performance Verification Guide
3
Frequency Response Test (PA On):
Reference
Value 1
Reference
Value 2
Signal
Generator
Output
Frequency
Power Meter
Measurement
Value A1
Spectrum
Analyzer
Measurement
Value A2
System
Error
[1]
Global
Error
[2]
Frequency
Response
[3]
Limit
Pass/Fail
100 kHz
100
kHz to
3.2
GHz:
<0.7 dB,
<0.3 dB
(typical)
3.2 GHz
to 6.5
GHz:
<0.9 dB,
<0.5 dB
(typical)
500 MHz
2 GHz
3 GHz
4 GHz
5 GHz
6 GHz
6.499995
GHz
Note
[1]
: System Error = Power Meter Measurement Value A1 - Reference Value 1
Note
[2]
: Global Error = Spectrum Analyzer Measurement Value A2 - Reference Value 2
Note
[3]
: Frequency Response = |Global Error - System Error|
Second Harmonic Distortion Test:
Output Frequency of
Signal Generator
Delta
Calculation Result
[1]
Limit
Pass/Fail
300 MHz
>45 dBm
1 GHz
Note
[1]
: Calculation Result = -20 dBm - Delta (-20 dBm is the level input into the mixer).
TOI Distortion Test:
Center Frequency
Delta
Calculation Result
[1]
Limit
Pass/Fail
299.8 MHz
>11 dBm, >+15
dBm (typical)
300.4 MHz
999.8 MHz
1.0004 GHz
Note
[1]
: Calculation Result = -20 dBm - Delta/2 (-20 dBm is the level input into the mixer).