EasyManuals Logo
Home>Rigol>Measuring Instruments>RSA5000 Series

Rigol RSA5000 Series Performance Verification Guide

Rigol RSA5000 Series
43 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #41 background imageLoading...
Page #41 background image
Appendix RIGOL
RSA5000 Performance Verification Guide
3
Frequency Response Test (PA On):
Reference
Value 1
Reference
Value 2
Signal
Generator
Output
Frequency
Power Meter
Measurement
Value A1
Spectrum
Analyzer
Measurement
Value A2
System
Error
[1]
Global
Error
[2]
Frequency
Response
[3]
Limit
Pass/Fail
100 kHz
100
kHz to
3.2
GHz:
<0.7 dB,
<0.3 dB
(typical)
3.2 GHz
to 6.5
GHz:
<0.9 dB,
<0.5 dB
(typical)
500 MHz
2 GHz
3 GHz
4 GHz
5 GHz
6 GHz
6.499995
GHz
Note
[1]
: System Error = Power Meter Measurement Value A1 - Reference Value 1
Note
[2]
: Global Error = Spectrum Analyzer Measurement Value A2 - Reference Value 2
Note
[3]
: Frequency Response = |Global Error - System Error|
Second Harmonic Distortion Test:
Output Frequency of
Signal Generator
Delta
Calculation Result
[1]
Limit
Pass/Fail
300 MHz
>45 dBm
1 GHz
Note
[1]
: Calculation Result = -20 dBm - Delta (-20 dBm is the level input into the mixer).
TOI Distortion Test:
Center Frequency
Delta
Calculation Result
[1]
Limit
Pass/Fail
299.8 MHz
>11 dBm, >+15
dBm (typical)
300.4 MHz
999.8 MHz
1.0004 GHz
Note
[1]
: Calculation Result = -20 dBm - Delta/2 (-20 dBm is the level input into the mixer).

Other manuals for Rigol RSA5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol RSA5000 Series and is the answer not in the manual?

Rigol RSA5000 Series Specifications

General IconGeneral
BrandRigol
ModelRSA5000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals