Appendix RIGOL
RSA5000 Performance Verification Guide
5
Note
[1]
: System Error = Measurement Value of the Power Meter - (+10 dBm - Attenuation of the Program-controlled
Attenuator) - System Error Reference Value
Note
[2]
: Global Error = Maximum Peak of the Spectrum Analyzer - (+10 dBm - Attenuation of the Program-controlled
Attenuator) - Global Error Reference Value
Note
[3]
: Uncertainty = |Global Error - System Error|
10 MHz Reference Output Accuracy Test:
Frequency Counter
Measurement Result
Output
Accuracy
[1]
Limit
Pass/Fail
<10 MHz × (1 ppm × time since last calibration + 0.5
ppm + 1 ppm)
Note
[1]
: Output Accuracy = |Frequency Counter Measurement Result 10 MHz|