Name Address Type R W Description
RTC Second 0157
H
uint8 A A Read/Write current second (0..59)
Zero scale 0300
H
execute N A Zero the scale. Returns:
0 for success
1 for operation canceled
2 for operation in progress
3 for scale is in error
4 for scale is over or under loaded
5 for ADC already busy
6 for scale is in motion
7 for outside of allowed (zero, tare etc.)
band
8 resolution too low
9 command not implemented
A duplicate point
B scale is in high resolution mode
C print ID has reached its maximum value
D current date and time could not be
determined
E parameter passed was not valid
F operation not allowed in trade mode.
Tare scale 0301
H
execute N A Tare the scale. Return value as described
for the zero scale register.
Preset tare 0302
H
execute N A Set preset tare value. Return value as
described for the zero scale register.
Gross/Net 0303
H
execute N A Set scale to gross or net. Argument values:
1 sets to Gross
2 sets to Net
all other values will toggle.
Return value as described for the zero
scale register.
High Resolution 0304
H
execute N A Set high resolution mode. Argument
values:
1 sets to high resolution mode
2 sets normal resolution mode
3 sets resolution to DB setting
all other values will toggle between high
and normal resolution.
Returns 1 for high resolution or 0 for
normal resolution.
Lua Status 2 0305
H
uint32 A N 32 bit number containing lua status bits as
follows:
0: Hi-Res
1-2: Dispmode values:
0: Display primary weight
1: Display pieces
2: Display secondary weight
3: Display tertiary weight
3-4: Range: 0 (range 1), 1 (range 2), 2
(range 3)
5: Menu active
8: USB Partition Notiļ¬cation
25: RTC changed
26: Weight changed
27: IO changed
83 C500-600-2.0.3