EasyManua.ls Logo

RKC INSTRUMENT PF Series - Page 82

Default Icon
170 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4. BASIC OPERATION
IMR02L04-E3
4-34
Continued from the previous page.
Name Symbol Data range
Factory set
value
Input type
(A continuance)
INP
24: Voltage (low) 0 to 1 V DC
25: Voltage (low) 10 to 10 mV DC
26: Voltage (low) 100 to 100 mV DC
27: Voltage (low) 1 to 1 V DC
Voltage (high) input group
17: Voltage (high) 0 to 10 V DC
18: Voltage (high) 0 to 5 V DC
19: Voltage (high) 1 to 5 V DC
20: Voltage (high) 5 to 5 V DC
21: Voltage (high) 10 to 10 V DC
Current input group
15: Current 0 to 20 mA DC
16: Current 4 to 20 mA DC
To switch Input group, refer to How to switch Input group
(P. 3-8).
Based on
model code.
When not
specifying: 0
Display unit
UNIT
0: C
1: F
0
Decimal point position
PGDP
0: No decimal place
1: One decimal place
2: Two decimal place
3: Three decimal place
4: Four decimal place
TC input: Only 0 or 1 can be set.
RTD input: From 0 to 2 can be set.
Voltage (V)/Current (I) input: From 0 to 4 can be set.
Based on
model code.
When not
specifying: 1
Input range high
PGSH
TC/RTD inputs:
Input range low to Maximum value of the selected
input range
Voltage (V)/Current (I) inputs:
19999 to 32000
Varies with the setting of the Decimal point position.
TC/RTD:
Maximum
value of the
selected input
range
V/I: 100.0
Input range low
PGSL
TC/RTD inputs:
Minimum value of the selected input range to Input
range high
Voltage (V)/Current (I) inputs:
19999 to 32000
Varies with the setting of the Decimal point position.
TC/RTD:
Minimum
value of the
selected input
range
V/I: 0.0
Input error determination point (high)
POV
Input range low
(5 % of Input span) to
Input range high (5 % of Input span)
Maximum setting value of Input error determination point
(high): 32767 (excluding decimal point)
Minimum setting value of Input error determination point
(low): 19999 (excluding decimal point)
Input range
high (5 % of
Input span)
Input error determination point (low)
PUN
Input range
low (5 % of
Input span)
Burnout direction
BOS
0: Upscale
1: Downscale
Valid only when the Voltage (low) input group selected.
0
Square root extraction
SQR
0: Unused
1: Used
0
Power supply frequency
PFRQ
0: 50 Hz
1: 60 Hz
0
Sampling cycle
SMP
0: 50 ms
1: 100 ms
2: 250 ms
1
PV bias *
PB
Input span to
Input span 0
PV digital filter *
DF
0.0 to 100.0 seconds
0.0: Unused
0.0
PV ratio *
PR
0.001 to 9.999 1.000
PV low input cut-off *
L-CUT
0.00 to 25.00 % of Input span 0.00
* It is not necessary to change the operation mode to the Reset mode (RESET) when setting the parameters. The parameters may