Beacon 410 Gas Monitor Operators Manual Viewing and Changing Channel Parameters 39
ALM-1 STROBE Setting
(NonResettable STROBE)
If set to Resettable STROBE, the strobe can be turned off with
the Reset Switch while an alarm 1 condition still exists.
If set to NonResettable STROBE, the strobe cannot be turned off
with the Reset Switch while an alarm 1 condition still exists.
If set to NO STROBE, the strobe will not activate due to an
alarm 1 condition.
* This parameter appears only if the STROBE Installation
parameter in the Global Menu is set to Strobe Installed
and the Strobe Setting parameter in the Configuration
Menu is set to USE w/ Alarms Only or USE w/ Alarms and
Fail.
Alarm-2 Level
(See the Beacon 410 Detector
Head Specification sheet for the
detector head installed on this
channel.)
The gas reading at which an alarm 2 condition occurs and the
Beacon 410 activates the alarm 2 circuit for this channel.
Alarm-2 Inc/Dec
(INCREASING)
Indicates if the alarm 2 condition is activated by gas readings
INCREASING or DECREASING to the alarm 2 level.
Alarm-2 Latching
(LATCHING)
If set as LATCHING, you must press the RESET button to reset
the alarm 2 circuit after the alarm 2 condition passes.
If set as SELF-RESETTING, the Beacon 410 automatically resets
the alarm 2 circuit after the alarm 2 condition passes.
Alarm-2 ON Delay
(1 Second)
The amount of time the Beacon 410 delays activation of the
alarm 2 circuit after the gas reading reaches the alarm 2 setpoint.
Alarm-2 OFF DELAY
(0 Seconds)
The amount of time the Beacon 410 delays resetting the alarm 2
circuit after the alarm 2 condition passes.
* This parameter appears only if the ALARM-2 Latching
parameter is set to SELF-RESETTING.
ALM-2 STROBE Setting
(NonResettable STROBE)
If set to Resettable STROBE, the strobe can be turned off with
the Reset Switch while an alarm 2 condition still exists.
If set to NonResettable STROBE, the strobe cannot be turned off
with the Reset Switch while an alarm 2 condition still exists.
If set to NO STROBE, the strobe will not activate due to
an alarm 2 condition.
* This parameter appears only if the STROBE Installation
parameter in the Global Menu is set to Strobe Installed
and the Strobe Setting parameter in the Configuration
Mode is set to USE w/ Alarms Only or USE w/ Alarms and
Fail.
Alarm-3 Level
(See the Beacon 410 Detector
Head Specification sheet for the
detector head installed on this
channel.)
The gas reading at which an alarm 3 condition occurs and the
Beacon 410 activates the alarm 3 circuit for this channel.
Alarm-3 Inc/Dec
(DECREASING for oxygen,
INCREASING for all other
channel types)
Indicates if the alarm 3 circuit is activated by gas readings
INCREASING or DECREASING to the alarm 3 level.
Table 7: Configuration Menu Parameters (Continued)
Channel Parameter
(Factory-Set Value)
Description