- 15 -
<Connection of Gas Sampling Probe and Gas Sampling Hose>
Attach the gas sampling probe to the end of the gas sampling hose.
Connect the sampling hose securely to the gas inlet (GAS IN) of the gas monitor.
Insert the sampling hose to the gas inlet (GAS IN) until it clicks into
place to ensure connection.
CAUTION
Use only a gas sampling hose specified by RKI.
Use the gas monitor with the gas sampling probe connected so that no foreign substance is
drawn into it.
Connect a gas sampling probe and a gas sampling hose by fastening them manually without
using any tool. If they are fastened too tightly using a tool, the plastic part of the gas sampling
probe may be broken.
CAUTION
Connecting tubing to the GAS OUT fitting suppresses the gas reading. RKI recommends that no
tubing be attached to the GAS OUT fitting. If the application requires routing of the exhaust gas,
keep the tubing length as short as possible.