Press o to mute Press o to confirm Maintenance message
M24 Occlusion:
The insulin cannot flow freely. Change the complete infusion
set. If the Maintenance message occurs again, change the
cartridge as well.
M25 Loantime over (loan pumps only):
The loan time of your pump has expired. Contact your local
pump support (see inside back cover) to discuss appropriate
action. After a Maintenance message M25 the pump will be in
STOP mode.
M26 Cartridge change not completed:
The cartridge change process was interrupted. Repeat the
cartridge change.
M27 Data download failed:
The data transfer to a computer has been interrupted. Restart
the data download.
M28 Pause mode timeout over:
Your pump has been in PAUSE mode for more than 15 minutes
and has switched to STOP. TBR and boluses have been
cancelled. Restart insulin delivery and reprogram TBR and
boluses as appropriate.