EasyManua.ls Logo

Roche Accu-Check Insight - General Safety and Product Symbols

Roche Accu-Check Insight
211 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
166
Appendix B: Configuration parameters
Parameter Standard parameter settings
Typical range/value available on
the pump
Maximum basal delivery (maximum hourly basal rate
combined with maximum temporary basal rate)
62.5 U/h 62.5 U/h
Hourly basal rates increment
0.01 (from 0.02U/h to 5.00 U/h) and
0.1 (from 5.00 U/h up to 25.0 U/h)
Hourly basal rates 0, 0.02–25.0 U/h 0, 0.02–25.0 U/h
Temporary Basal Rate
TBR increases or decreases in increments of
TBR duration
TBR duration increments
100 %
10 %
2 h
15 min
0–250 %
10 %
15 min–24 h
15 min
Basal profiles available 5 5
Bolus amount per bolus delivery 0–25.0 U 0–25.0 U
Bolus amount increments for Quick Bolus 0.5 U 0.1, 0.2, 0.5, 1.0, 2.0 U
Bolus amount increments for Standard Bolus, Extended
Bolus and Multiwave Bolus
0.05 U (0–2 U)
0.1 U (2–5 U)
0.2 U (5–10 U)
0.5 U (10–20 U)
1.0 U (20–25 U)
Bolus duration (Extended Bolus, Multiwave Bolus) 2 h 15 min to max. 24 h

Table of Contents

Other manuals for Roche Accu-Check Insight

Related product manuals