166
Appendix B: Configuration parameters
Parameter Standard parameter settings
Typical range/value available on
the pump
Maximum basal delivery (maximum hourly basal rate
combined with maximum temporary basal rate)
62.5 U/h 62.5 U/h
Hourly basal rates increment
0.01 (from 0.02U/h to 5.00 U/h) and
0.1 (from 5.00 U/h up to 25.0 U/h)
Hourly basal rates 0, 0.02–25.0 U/h 0, 0.02–25.0 U/h
Temporary Basal Rate
TBR increases or decreases in increments of
TBR duration
TBR duration increments
100 %
10 %
2 h
15 min
0–250 %
10 %
15 min–24 h
15 min
Basal profiles available 5 5
Bolus amount per bolus delivery 0–25.0 U 0–25.0 U
Bolus amount increments for Quick Bolus 0.5 U 0.1, 0.2, 0.5, 1.0, 2.0 U
Bolus amount increments for Standard Bolus, Extended
Bolus and Multiwave Bolus
0.05 U (0–2 U)
0.1 U (2–5 U)
0.2 U (5–10 U)
0.5 U (10–20 U)
1.0 U (20–25 U)
Bolus duration (Extended Bolus, Multiwave Bolus) 2 h 15 min to max. 24 h