191
Lag Time With the Lag Time you can specify a
delay between programming a bolus and
the actual beginning of the bolus
delivery.
Maintenance A Maintenance message will be
triggered when the pump requires your
immediate action to maintain its
functionality. After a Maintenance
message the pump will be in PAUSE or
STOP mode. Restart insulin delivery as
needed.
Modes With Modes, you can pre-define the way
your pump signals messages in dierent
life situations. By default you can adjust
the modes for Normal, Vibrate, Quiet and
Loud.
Occlusion An occlusion is a blockage which
prevents the insulin from flowing
correctly from the insulin pump into the
body.
PAUSE When you open the adapter and after
some Maintenance messages, your
pump switches to PAUSE mode and
interrupts the insulin delivery. However,
if you close the adapter within
15 minutes, the insulin delivery will
restart after confirmation, and the
missing amount of basal insulin will be
delivered additionally as well as boluses,
and TBRs will continue. If the pump is in
PAUSE mode for more than 15 minutes,
it will switch to STOP mode.