Electromagnetic fields and hazardous areas
w WARNING
Do not use your pump near electromagnetic fields such as
radar or antenna installations, high-voltage sources, X-Ray
sources, computer tomography, CAT scan and MRI. These and
all other sources of electrical current may cause your pump to
fail. Insulin delivery may stop and the Error E7 occurs.
Always stop your pump, remove it from your body, and leave
it outside these areas. See Appendix A: Technical Data, page151
for more information about electromagnetic fields.
Your pump has been tested against, and complies with, the
regulations on electromagnetic interference. Security systems in
airports and anti-theft monitoring devices, such as those in
department stores, should not aect the functioning of your
pump.
There are many devices that emit electromagnetic radiation, for
example mobile phones. It cannot be completely ruled out that
one of such devices could aect your pump. It is therefore
recommended that you keep your pump at least 10 cm
(4 inches) away from these devices while they are active.