Meter Memory, Setup, and Data Transfer
23
3
Press and release to
turn the meter on.
The flashing test strip
symbol appears on the
display.
Press and hold
until
set‑up appears on the
display.
1 2
Press and release
repeatedly until the bell
symbol, OFF, set‑up,
and the flashing A‑1
appear on the display.
Press and release
or
to switch between On
and OFF.
Press and release
to
set your choice.
If On is selected, the
hour flashes.
3
Press and release or
to adjust the hour.
Press and release
to
set the hour.
The minutes flash.
4
NOTE
Test reminders:
• beep at the same time each day.
• beep every 2minutes up to 3times.
• turn o by inserting a test strip or pressing any
button.
• If a test was performed within 15minutes of a
test reminder, the test reminder does not beep.
• If the meter is on at the test reminder time, the
test reminder does not beep.
• Exposure to cold conditions may disable test
reminders until the meter is turned on.
• If a test reminder is o in set‑up mode, any
subsequent test reminders are also o. For
example, if you set A‑1 but turn o A‑2, then
A‑3 and A‑4 will automatically be o.
Setting the Test Reminders
79718_07255233050_01_EN.indb 23 5/22/2019 1:44:54 PM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Accu-Chek Performa Nano and is the answer not in the manual?

Roche Accu-Chek Performa Nano Specifications

General IconGeneral
BrandRoche
ModelAccu-Chek Performa Nano
CategoryBlood Glucose Meter
LanguageEnglish

Related product manuals