Roche/Hitachi cobas c 311 analyzer Record structure
Comment record, that follows the result record (result data flag)
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 E-17
Comment record, that follows the result record (result data flag)
cobas type (Upload)
C|1|I|16|l[CR]
(1)(2)(3)(4)(5)
Comments
No
Field
No.
Field Type Max
cobas Type
(1) 1 Record Type ID ST 1 ‘C’ fixed.
(2) 2 Sequence Number NM 6 Indicates sequence No. Normally it is ‘1’
(3) 3 Comment Source ST 1 Comment is send from analyzer. "I" is displayed.
(4) 4 Comment Text NM 3
(5) 5 Comment Type ST 1 ‘I’ fixed.
Note: Skip this record when no data alarm for measured value.
Note: See Table E-3 for data alarm No. and alarm message.