Roche/Hitachi cobas c 311 analyzer Record structure
List of result data flags
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 E-22
List of result data flags
Table E-3 Data Alarm List
Alarm
code
Alarm Result
Data
Photo-
metry
ISE Printer Screen
Host
no alarm 0
1 ADC abnormal X X ADC.E ADC.E 1
2 Cell blank abnormal X >Cuvet >Cuvet 2
3 Sample short SPACE* X X Samp.S Samp.S 3
4 Reagent short SPACE* X X Reag.S Reag.S 4
5 ABS over X >Abs >Abs 5
6 Prozone error X >Proz >Proz 6
7 Reaction limit over(all photometry points) X >React >React 7
8 Reaction limit over(second and subsequent
photometry points)
X >React >React 8
9 Reaction limit over(third and fourth and subsequent
photometry points)
X >React >React 9
10 Linearity abnormal(nine or more) X >Lin >Lin 10
11 Linearity abnormal(eight or less) X >Lin >Lin 11
12 Standard Solution 1 Absorbance Abnormal X S1A.E S1A.E 12
13 Duplicate Error X Dup.E Dup.E 13
14 STD Error X X Std.E Std.E 14
15 Sensitivity Error X Sens.E Sens.E 15
16 Calibration Error X X Cal.E Cal.E 16
17 Convergence Error X SD.E SD.E 17
18 Noise error X ISE.N ISE.N 18
19 Level error X ISE.E ISE.E 19
20 Slope Error X Slop.E Slop.E 20
21 ISE Slope Marginal X Prep.E Prep.E 21
22 Internal Concentration Abnormal X Istd.E IStd.E 22
23 Sample range over X <>Test <>Test 23
24 Data error in Compensate test X X CmpT.E CmpT.E 24
25 Unable to calculate compensate test SPACE X X CmpT.? CmpT.? 25
26 PANIC value over (upper) X X >Test >Test 26
27 PANIC value over (lower) X X <Test <Test 27
28 Random Error [R-4s] X X R4SD R4SD 28
29 System Error 1 [2-2sA] X X S2-2Sa S2-2Sa 29
30 System Error 2 [2-2sW] X X S2-2Sw S2-2Sw 30
31 System Error 3 [4-1sA] X X S4-1Sa S4-1Sa 31
32 System Error 4 [4-1sW] X X S4-1Sw S4-1Sw 32
33 System Error 5 [10xA] X X S10Xa S10Xa 33
34 System Error 6 [10xW] X X S10Xw S10Xw 34
35 QC Error 1 X X Q3SD Q3SD 35
36 QC Error 2 X X Q2.5SD Q2.5SD 36
37 Calculated test error X X ClcT.E ClcT.E 37
38 Overflow SPACE X X Over.E Over.E 38
39 Calculation not possible SPACE X X Calc.? Calc.? 39
40 Outside of expected value(upper) X X H n.a.
41 Outside of expected value(lower) X X L n.a.
42 Edited test X X Edited Edited 42
43 Calibration result abnormal SPACE* X X Cal.E Cal.E 43
44 Repeat limit over(upper) X X >Rept >Rept 44
45 Repeat limit over(lower) X X <Rept <Rept 45
46 ABS maximum over SPACE X Samp.? Samp.? 46
51 Response Error 1 X Rsp1.E Rsp1.E 51
52 Response Error 2 X Rsp2.E Rsp2.E 52
53 Conditioning Error X Cond.E Cond.E 53
55 ISE Range over X >ISE >ISE 55
56 Kinetic unstable (Prozone error 2) X >Kin >Kin 56
57 Serum index 1 interference X X >Index >Index 57
58 Serum index multiple interference X X >Index >Index 59
59 Mixing abnormal X X Mix.E Mix.E 59
60 Drop in ultrasonic output X X <Mix <Mix 60
71 Detergent short X Det.S Det.S 71
72 Sample Clot SPACE X Samp.C Samp.C 72
101 Expired reagent X ReagEx ReagEx 101