Roche/Hitachi cobas c 311 analyzer Communication trace
Real Time Test Results
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 F-3
Real Time Test Results
(1) When the result value is within normal range.
cobas c 311 analyzer sends test result of sample ID=000004,
sequence No.=40, and Position=5.
H|\^&|||cobas c 311^1|||||host|RSUPL^REAL|P|1[CR]
P|1 [CR]
O|1| 000004|40^50005^005^^S1^SC|^^^10^\^^^30^3\^^^40^|
R||||||N ||||1|||||||20051220095504|||F[CR]
C|1|I| ^
^ ^ ^ |G
[CR]
R|1|^^^10/|1.25|ulU/ml||N||F||admin|||P1[CR]
C|1|I|0|I[CR]
R|2|^^^30/2|0.091|ug/dL||N||F||admin|||P1[CR]
C|1|I|0|I[CR]
R|3|^^^40/inc|1.17|ng/mL||N||F||admin|||P1[CR]
C|1|I|0|I[CR]
L|1|N[CR]
(2) When the result value is less than normal range.
cobas c 311 analyzer sends a test result of sample
ID=000002, sequence No.=3, and Position=2.
H|\^&|||cobas c 311^1|||||host|RSUPL^REAL|P|1[CR]
P|1 [CR]
O|1| 000002|3^50002^002^^S1^SC|^^^10^|R||||||N||||1|||||||
20051220104418|||F[CR]
R|1|^^^10/|0.163|mlU/ml||L||F||admin|||P1[CR]
C|1|I|45|I[CR]
L|1|N[CR]
(3) When the result value is a qualitative test.
cobas c 311 analyzer sends a test result of sample
ID=000010, sequence No.=442, and Position=1.
H|\^&|||cobas c 311^1|||||host|RSUPL^REAL|P|1[CR]
P|1
[CR]
O|1| 000010|442^50001^001^^S1^SC|^^^672^|R||||||
N||||1|||||||
20051220104418|||F
[CR]
R|1|^^^400/|-1^0.303|umol/l||N||F||admin|||P1[CR]
C|1|I|45|I[CR]
L|1|N
[CR]