Roche/Hitachi cobas c 311 analyzer Appendix A - Instrument Alarms
Instrument Alarm List
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 G-2
Appendix A - Instrument Alarms
Instrument Alarm List
Alarm Type Alarm No
Alarm Message Alarm Description Alarm Remedy
110 001 Abnormal Receiving Text
from Host
Abnormal Text has been received from Host. a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
111 002 Abnormal TS from Host No response for the inquiry within a set time in
GUI: [Utility]-[System]-[Host Communication].
a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
Resume operation; if alarm recurs, call Technical Support.
111 003 Abnormal TS from Host No response for the inquiry within 10 minutes. a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
111 005 Abnormal TS from Host Application code requested by the host is not
registered in the analyzer; or received
application code other than Photometric test,
ISE test or serum index test.
a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
111 007 Abnormal TS from Host Received request for increased or decreased
quantity for other than Sample Type 2 (Urine)
of ISE test.
a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
111 008 Abnormal TS from Host Received request for ISE test is wrong. a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.