Roche/Hitachi cobas c 311 analyzer Appendix A - Instrument Alarms
Instrument Alarm List
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 G-3
Alarm Type Alarm No
Alarm Message Alarm Description Alarm Remedy
111 009 Abnormal TS from Host Received request for serum index test is
wrong.
a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
111 010 Abnormal TS from Host Received comment even though the mode was
set for no transmission of comments.
a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
112 002 Abnormal Automatic
Rerun TS from Host
No response for the inquiry within a set time in
GUI: [Utility]-[System]-[Host Communication].
a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
112 003 Abnormal Automatic
Rerun TS from Host
No response for the inquiry within 10 minutes. a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
112 005 Abnormal Automatic
Rerun TS from Host
Application code requested by the host is not
registered in the analyzer; or received
application code other than Photometric test,
ISE test or serum index test.
a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
112 007 Abnormal Automatic
Rerun TS from Host
Received request for increased or decreased
quantity for other than Sample Type 2 (Urine)
of ISE test.
a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
112 008 Abnormal Automatic
Rerun TS from Host
Received request for ISE test is wrong.
Received request for other than pair of Na, K
or Na, K, Cl for ISE test.
a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
112 009 Abnormal Automatic
Rerun TS from Host
Received request for serum index test is
wrong. Received request for increased or
decreased quantity.
a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.