Roche/Hitachi cobas c 311 analyzer Communication trace
Batch Test Selection Information
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 F-4
(4) When it is a control sample.
H|\^&|||cobas c 311^1|||||host|RSUPL^REAL|P|1[CR]
P|1
[CR]
O|1| 17222200 |10096^30085^085^^QC^SC|^^^672^|||||||Q||||1
|||||||20051220104418|||F[CR]C|1|I|^^
^^|G
[CR]R|1|^^^10/|1.26|ulU/mL||L||F||admin|||P1[CR]
C|1|I|45|I[CR]
L|1|N[CR]
Batch Test Selection Information
Download test selection information of sample ID=000051
from HOST.
H|\^&|||host^1|||||cobas c 311|TSDWN^BATCH|P|1[CR]
P|1 [CR]
O|1| 000051|^^^^S1^SC|^^^10^/^^^30^
3/^^^40^|R||||||A||||1
||||||||||O
[CR]
C|1|I| ^
^ ^ ^ |G[CR]
L|1|N[CR]