28 Warning messages
Evaporation of samples or reagents Evaporation of samples or reagents may lead to incorrect
or invalid results. The term reagent applies to all reagent
cassettes and system reagents.
r Sample material may evaporate if left open. Do not
leave samples open for any length of time.
r Open the cover of the reagent disk only when
necessary.
r Reagents may evaporate if the cover of the reagent
disk is left open. Do not leave the cover of the reagent
disk open longer than necessary.
r Do not use improperly stored reagents. Ensure that
reagents are stored according to the Instructions for
Use.
r Do not use expired reagents.
Incorrect reagent volume Incorrect reagent handling may cause an undetectable
loss of reagent. The term reagent applies to all reagent
cassettes and system reagents.
r Always store reagents according to the specified
storage conditions as stated in the Instructions for Use
for the test.
r Do not use a reagent cassette or reagent bottle whose
reagent has spilled.
Expired calibrators or controls Use of expired calibrators or controls may lead to
incorrect or invalid results.
r Calibrators and controls may evaporate if left open. Do
not leave calibrators or controls open for any length of
time.
r Do not use improperly stored calibrators or controls.
Ensure that calibrators and controls are stored
according to the Instructions for Use.
r Do not use expired calibrators or controls.
Overfilling of sample containers Overfilling of sample containers may cause splashing of
the sample during transportation.
r Leave approximately 10 mm clearance between the
sample liquid and the top of the tube or cup.