cobas c 513 analyzer · Software Version 02-05 · Safety Manual · 6.0
29Warning messages
Incorrect handling of samples Incorrect handling of samples may cause incorrect
results.
r Do not load samples on the analyzer that have an
unusually high viscosity.
r Before you load samples on the analyzer, wait for the
samples to be at room temperature.
Incorrect use of barcode labels Incorrect use of barcode labels may cause incorrect
results.
r Use barcode labels of the correct specification.
r When you attach barcode labels to sample tubes,
ensure they are readable.
r To prevent undetected barcode read errors, use
barcodes with check digits.
r Ensure that all ITF barcode labels used have the same
number of digits.
q For more information about barcode labels, see the
list of barcode specifications in the Operator’s Manual or
User Assistance.
Incorrect loading of sample containers on
racks
Incorrect loading of sample containers may cause
incorrect results.
r Load sample containers of the correct specification on
racks.
r Specify rack ranges for sample types. Load sample
containers and types in the correct rack ranges.
r Only use sample tube types on CTS racks that are
configured on the system.
r Load sample tube types in the correct rack ranges.
q For more information about sample containers, see
the list of sample container specifications in the
Operator’s Manual or User Assistance.