.1
15 / 93
The analyzer will send inquiries for test and rerun orders only
when this function is enabled.
If disabled, test orders have to be made either manually by the
operator at the analyzer or by download from the host.
Query Timeout
(10 - 300)
*10 sec
Time before cobas c 111 instrument assumes that the host
does not answer.
This function allows the operator to set a timeout interval for
‘Query’ for Routine, STAT and Rerun Samples. The specified
timeout interval determines how long the analyzer will wait for a
response from the host. The range is 10-300 seconds, with a
default of 10 seconds.
Note: This function cannot be disabled.
Send Result Mode
OFF
*COMPLETE
IMMEDIATE
Allows Real-Time Communications for results
No results are sent automatically
Send results only if all results are complete avail.
Send each result as soon as available
Note:
QC results will be sent immediately if not switched OFF,
QC results can not be sent grouped by QC sample.
Only when this function is enabled, the analyzer will send
analytical result records to the host computer according to the
description above.
If disabled, result records have to be sent manually to the host
computer or the host may send an inquiry to upload such results.
Raw-Data
*OFF
ON
Raw data records are not sent
Raw data records (M.RR) are added to result record (if raw
data is still available)
Trace
ON
*OFF
A communication trace file is written
No communication trace file is written
When this function is enabled, the content of the communication
with the host can be stored at the analyzer. The Communication
Trace report can be saved via media interface to a Memory Stick.
This report can be used as an analysis tool if a problem occurs.
System ID
char[10]
*c111
The instrument device identification
Name used on reports and in communications for the cobas c
111 instrument system or the ID number used to identify the
system in communications with the host.