Field Name Reference Format
Comments (Data Content) H|7.1.
Comment or Special
Instructions
07.1.11 CM A comment regarding the message is entered here.
cobas c 111 instrument transmits data:
To indicate the classification of the transmitted message from the analyzer
one of the defined messages (see above) is transmitted.
The first component element indicates the purpose for the message, e.g.
TSREQ: Test request inquiry
RSUPL: Result report
The second component element indicates the cause of the message.
REAL: Message automatically generated
BATCH: Message generated due to operator command
REPLY: Response to an inquiry from a host
Or in case of HL7 based Manufacturer spec. recs.:
INU^U05: Automated equipment inventory update (MSR)
cobas c 111 instrument receives data:
To indicate classification of the transmitted message from the host, one of
the defined messages (see above) shall be sent by the host.
Processing ID 07.1.12 ST Indicates the processing method for the messages. Currently ‘P’ is used.
Version No. 07.1.13 NM Enter version number of the communication program. Currently ‘1’ is used.
Date and Time of
Message
07.1.14 TS Format: YYYYMMDDHHMMSS.
EndOfRecord

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cobas C111 and is the answer not in the manual?

Roche Cobas C111 Specifications

General IconGeneral
BrandRoche
ModelCobas C111
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals