Field Name Reference Format
Comments (Data Content) M.CR
result states M.CR.12 CM A – Accepted (with or without flags)
cobas c 111 instrument transmits data:
A^<operator> the user who accepted the cal will be submitted
O – Calculated no flags
F – Calculated with flags
X – cannot be measured (less than calculated)
Result characteristics M.CR.13 Not used / ignored by cobas c 111 instrument
Curve Parameters
M.CR.14 CM list of test number specific (application) curve parameters, e.g., A, B, C, D
for LogitLog4
NOTE: the parameter transmitted is the original internal data. For linear
models the factor is usually displayed as 1/parameter
curve parameter[0]^
curve parameter[1]^
curve parameter[2]^
curve parameter[3]^
curve parameter[4]
cobas c 111 instrument transmits data: Scientific notation i.e. 1.3454E2
ISE: slope^solution factor Example Page 55
Calibrators Container
Description
M.CR.15 CM field components:
carrier type^^^calibrator lot number
cobas c 111 instrument transmits data: only one field for a calibrator
(set)
carrier type = “SD”
calib. lot nr (8 chars)
Calibrator Result
text
{^ ... }
{\...}
M.CR.16 CM
\CM
..
effective signal (only one value per single measurements, n values for
replicates); this field component is repeated for each calibrator level
Nbr of replicates is in M.CR.10
cobas c 111 instrument transmits data:
Repeated from STD1 to max. STD6 for as many as there are. Format
<Targetvalue> (Conc)
^<Rate – dAbs or dAbs/Min>
^<Replicate1 – dAbs or dAbs/Min> (Rate)
^<Replicate2 – dAbs or dAbs/Min> (Rate)
^<Data Alarm> i.e. highest priority flag of a STD measurement
^<Prozone Value> (Antigen Excess Rate)
ISE: In case of an ISE calibration, the Prozone Value contains the ISE
Standard Signal (as opposed to Sample Signal in Rate)
Example Page 55
EndOfRecord