41 <RR
Outside of expected value (lower)
Description
For patient samples, the calculated concentration is less than the lower limit of the
expected value range.
nnnnn is the actual value which was checked against
46 CALC RANGE
Absorbance Maximum Over (nonLin curve)
Description
The absorbance of a sample is found equal or greater than the maximum theoretical
concentration. The result field will be left blank on the report and Data Review screen. This
blank result is transmitted, together with the alarm code ">" to the Host.
56
High Act
Low Act
Kinetic unstable (subst. Hitachi Rate Prozone check)
Description
The Absorbance change during measurement exceeds the accepted limit, i.e. the reaction
has taken off too fast. Possibility of substrate exhaustion.
Or the absorbance change during the measurement is lower than the defined limit.
18 Ise Unstab
ISE Noise error
Description
The signal from the electrode(s) was not stable during the measurement.
19
Air Fluid
Seg Fluid
ISE Voltage Level error
Description
The length of the fluid segment was too short.
The end of the fluid segment was not detected. It takes too long for a sample segment to
reach the electrodes.
91
Air Isecal
Seg Isecal
Description
The length of the calibrator segment was too short.
The end of the calibrator segment was not detected. It takes too long for a calibrator
segment to reach the electrodes.
68 No Fluid
Description
The fluid sensors (1 or 2) could not detect any liquid.
90
No Isecal
Description
Fluid sensor 2 could not detect a calibrator.