ROCON LLC
11. To reassemble: Spring fits into the top of the plunger. The plunger fits inside the guide with the
spring going in first. The plunger seat fits on top of the orifice seat. With plunger in place, screw
the screw piece into the valve cover.
12. With the valve cover complete, set the pressure spring on top of the diaphragm and the valve
cover on top of the pressure spring. Install the 4 socket head screws.
13. With the spring clip pulled, slide the coil over the guide. When complete, push the spring clip over
the coil.
14. Insert the DIN connector into the coil.
15. Energize coil to confirm operation. If there is a problem, consult Service Department at Rocon
LLC.
Override Stem Broken
Spare Part: # 2705-CAP
If the override stem is broken, replace the control cover. An indication of a broken stem is that the valve
will not shut off in fault (or when power is removed), while the stem is in Normal position. This is
determined when the valve is disassembled and no foreign object was found that would force open the
diaphragm.
Check Valve/Remove Piston
Tools Required: 1 3/16” box wrench to remove from unit, retainer ring puller, ¼ inch flat nose screwdriver
Spare Part: Check Valve #2081
1. Remove the retainer ring.
2. Gently push the piston (water flow OUT side) with the screwdriver until dislodged. The O-ring
remains in place. Only the spring and the piston are removed. Clean with soap and water.
3. Check if the O-ring needs replacement.
4. Insert the spring over the piston and insert them into return port.
5. Replace the retainer ring.
6. Test to confirm that the unit is operational. If there are problems, contact the Service Department
at Rocon, LLC.
Cleaning
These meters do not require any special cleaning of the external surfaces. If cleaning is deemed
necessary, strong solvents, detergents, or chemicals should not be used. A damp cloth may be used to
wipe off dirt or debris.
Note
If used outside the parameters specified in this manual, the proper operation of the flowmeter cannot be
guaranteed.