DeltaPoint Uno Elbow Test Results
DELTA POINT UNO WATER FLOWMETER TEST
With and Without 90 Degree Fitting Installed Inlet Port
2.00
A
Master Flow Meter
2.00
A
Master Flow Meter
2.00
DeltaPoint Uno
B
C
Based on the above results, there is no difference in performance in the
installation of a 90 degree fitting in the inlet port of the flow switch on the
DeltaPoint Jr unit.
A = Master
GPM
B = DP Uno
Increasing
B = DP Uno
Decreasing
C = DP Uno
Increasing
C = DP Uno
Decreasing
1 0.98 0.99 0.98 0.95
2 1.97 1.98 1.98 1.99
3 2.96 2.96 2.96 2.98
4 3.96 3.95 3.98 3.99