EasyManua.ls Logo

Rohde & Schwarz FSH-Z2 - Replacing the Coax Cable Assembly

Rohde & Schwarz FSH-Z2
21 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Spare Part Replacement R&S FSH-Z2
1145.5767.02 3.8 E-1
Replacing the coax cable assembly
(See exploded view for references).
½ Open the instrument as described earlier.
½
Fig. 3-10 Connect the coax assy (item 10) to the N-SMA connector (item 07) using a torque wrench
½
Fig. 3-11 Connect the coax assy (item 10) to the unit using a torque wrench.
½ Bend the coax assy as shown in Fig. 3-5
½ Close the instrument as described earlier.