R&S FSH-Z2 Test Instructions
1145.5767.02 1.1 E-2
1 Performance Test
Test Instructions
• Values specified in the following sections are not guaranteed. Only the technical specifications
provided on the data sheet are binding.
• The values specified in the data sheet are the guaranteed limits.
• All non used ports are supposed to be terminated with a 50 Ohm load.
• Inputs for settings during measurements are shown as following:
[<KEY>] Press a key on the front panel, eg [SPAN].
[<SOFTKEY>] Press a softkey, e.g. [MARKER -> PEAK].
[<nn unit>] Enter a value and terminate by entering the unit, e.g. [12 kHz].
Successive entries are separated by [:], e.g. [ BW : MANUAL RES BW : 3 kHz ].
Measuring Equipment and Accessories
Item Type of
equipment
Specifications recommended Equipment
recommended
R&S order
no.
Use
1 Spectrum
Analyzer with TG
output
Frequency: 10 MHz to 3 GHz R&S FSH 1145.5850.13 Directivity
InsertionLoss
ReturnLoss
2 SWR Bridge Directivity > 40 dB , RL testport >
20 dB
R&S ZRB 2 0373.9017.53 ReturnLoss
3 Short Circuit
Termination
10 MHz to 3 GHz
N-type male
R&S FSH-Z30 1145.5773.02 Directivity
ReturnLoss
4
50-Ω termination
N-male
10 MHz to 3 GHz
Return loss ≤ -35 dB
R&S FSH-Z31 1145.5780.02 Directivity
InsertionLoss
ReturnLoss
5 Precision
Adapter
N-male to N-male Huber&Suhner:
32 N-50-0-51
ReturnLoss
6
50-Ω termination
N-female (2x)
10 MHz to 3 GHz Pasternack:
PE6032
ReturnLoss
7 Adapter N-male to N-male Huber&Suhner:
32 N-50-0-1
ReturnLoss
8 Cable Flexible cable RG213/U with N-
male connector on both sides.
L=approx 30 cm
Pasternack:
PE3496-12
InsertionLoss
ReturnLoss