Performance Test R&S FSH-Z2
1145.5767.02 1.2 E-2
Performance Test
Checking the directivity
Test equipment:
- Spectrum Analyzer (refer to section "Measurement Equipment and
Accessories", item 1)
-
50 Ohm termination (refer to section "Measurement Equipment and
Accessories", item 4)
- Short Circuit termination (refer to section "Measurement Equipment and
Accessories", item 3)
Test setup:
¾ Mount the R&S FSH-Z2 on to the spectrum analyzer (item 1).
Spectrum Analyzer settings:
- [ PRESET]
- [ FREQ ]
- [ START FREQ : 10 MHz ]
- [ STOP FREQ : 3 GHz ]
- [ BW : MANUAL VIDEO BW : 3 kHz ]
- [ MEAS : TRACKING GEN : MEASURE ]
- [ REFLECT CAL : CONTINUE ]
Measurement: ¾ Connect the Short Circuit termination (item 3) to the "TEST" Port of the
R&S FSH-Z2
¾ Press softkey [ CONTINUE ] on the Spectrum Analyzer.
¾ Replace the Short Circuit termination (item 3) on the "TEST" Port of
the R&S FSH-Z2 with the 50 Ohm termination (item 4).
¾
The curve shows the directivity (limited by returnloss of the used 50
Ohm termination, item 4).
Hint: ¾ As far as the datasheet states only typical values, the test limits are
some dB worse than these numbers!