EasyManua.ls Logo

Rohde & Schwarz NRP-Z51 - Calibration:test:deviation; Calibration:test:reference; Calibration:zero:auto[?] off | on | ONCE

Rohde & Schwarz NRP-Z51
83 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control - Commands R&S NRP-Z5x
1137.0463.12 6.4 E-7
By invoking the CALibration:TEST? command, the power difference in W between the test heater
s
witched on and off is determined and output.
CALibration:TEST:DEViation?
With CALibration:TEST:DEViation?, the power difference in W between the test heater switched on and
off is determined. The relative deviation of the power difference from the reference value stored during
calibration is output.
The CALibration:TEST:DEViation? query internally calls up the CALibration:TEST? and
CALibration:TEST:REFerence? queries. The relative deviation of the first value relative to the reference
value is calculated as follows:
1=
erence
n:TEST:REF
CALibratio
n:TESTCALibratio
iation n:TEST:DEVCALibratio
CALibration:TEST:REFerence?
During the calibration process, the power difference in W between the test heater switched on and off is
determined and stored in the calibration data set. CALibration:TEST:REFerence? yields the value
stored at the time of calibration.
CALibration:ZERO:AUTO[?] OFF | ON | ONCE
The commands CALibration:ZERO:AUTO ON and CALibration:ZERO:AUTO ONCE zeroes the three
measurement paths of the sensor. For this purpose, the test signal must be deactivated or the sensor
disconnected from the signal source. The sensor automatically detects the presence of any significant
power to be measured. This causes zeroing to be aborted and error message NRPERROR_CALZERO
to be output. The CALibration:ZERO:AUTO OFF is ignored. Zeroing takes four seconds at a minimum,
but at least as long as the selected averaging filter needs for settling (only fixed-filter mode).
Repeat zeroing
during warm-up after switching on or connecting the instrument
after a substantial variation of the ambient temperature
after fastening the sensor to an RF connector at high temperature
after several hours of operation
when very low-power signals are to be measured, e.g. less than 10
dB above the
lower measurement limit.
For zeroing switch off the test signal and do not remove the sensor from the signal source.
Apart from kee
ping the thermal balance, this has the advantage that the noise
superimposed on the test signal (e.g. from a broadband amplifier) can be detected on
zeroing and does not impair the measurement result.

Table of Contents

Related product manuals