Improving Level Performance
R&S
®
SMA100B
264User Manual 1178.3834.02 ─ 05
17. Close the "Trace/Pulse Data Notifications/ Trigger" dialog.
18. Press "Start Cont" to start the measurement.
7.4.4.12 How to Configure the Power Analysis Diagram
1. Select "Clk Syn/Power Sens" > "NRP Power Analysis".
2. Select "Configure" > "Configure Diagram".
3. Set "Max Level = 30dBm".
4. Set "Min Level = -40dBm".
5. Select "Auto Scale > Expand Only (Coarse)".
6. Enable "Show Grid".
7. Set "Background Color > Black."
7.5 How to Calibrate the Power Level with an R&S NRP
Power Sensor
Using a R&S NRP power sensor and the user correction function of the instrument,
you can compensate external frequency responses. The R&S SMA100B utilizes the
readings of the power sensor and creates a correction value table for controlling the
output level during operation.
To connect the R&S NRP to the R&S SMA100B
1. Connect the power sensor.
See Chapter 7.4.1, "Connecting R&S NRP Power Sensors to the R&S SMA100B",
on page 211.
2. Select "Clk Syn/Power Sens" > "NRP Sensor Mapping".
3. If the sensor is not detected automatically, select "Scan > Start".
The instrument scans the network and the USB connections for connected sensors
and lists all detected R&S NRP sensors in the mapping table.
4. In the "Mapping" column, assign the sensor to a sensor channel (see Chap-
ter 7.4.2, "NRP Sensor Mapping", on page 212
5. Close the dialog.
How to Calibrate the Power Level with an R&S NRP Power Sensor