Remote Control Commands
R&S
®
SMA100B
466User Manual 1178.3834.02 ─ 05
Usage: Event
Manual operation: See "Save To Ref" on page 228
:SENSe[:POWer]:SWEep:FREQuency:REFerence:DATA:POINts?
Queries the number of points from the reference curve in "Frequency" measurement.
Return values:
<Points> integer
Range: 10 to 1000
Example:
SENS:POW:SWE:FREQ:REF:DATA:POIN?
Queries the number of points from the reference curve in fre-
quency mode.
Usage: Query only
:SENSe[:POWer]:SWEep:FREQuency:REFerence:DATA:XVALues <XValues>
Sets or queries the x values of the two reference points, i.e. "Frequency X (Point A)"
and "Frequency X (Point B)" in "Frequency" measurement.
Parameters:
<XValues> string
Example:
SENSe:POW:SWE:FREQ:REF:DATA:XVAL 100MHZ,22GHZ
Sets the x value of reference "Point A" to 10 MHz, and the value
of "Point B" to 10 GHz.
Manual operation: See "Freq (X) / Pow (Y)" on page 228
:SENSe[:POWer]:SWEep:FREQuency:REFerence:DATA:YVALues <YValues>
Sets or queries the y values of the two reference points, i.e."Pow Y (Point A)" and
"Power Y (Point B)" in "Frequency" measurement.
Parameters:
<YValues> string
Example:
SENSe:POW:SWE:FREQ:REF:DATA:YVAL -10,25
Sets the y value of reference "Point A" to -10 dBm, and the
value of "Point B" to -25 dBm.
Manual operation: See "Freq (X) / Pow (Y)" on page 228
:SENSe[:POWer]:SWEep:FREQuency:RMODe <RMode>
Selects single or continuous mode for measurement mode frequency in power analy-
sis.
Power Sensor Measurement Subsystems