Waveform setup
R&S
®
RTA4000
70User Manual 1335.7898.02 ─ 08
Figure 5-6: Probe settings of R&S
RT-ZM30 with ProbeMeter measurement
Offset
Same as "Offset" in the "Vertical" menu, see "Offset" on page 57.
Copy to offset
Same as "Copy to offset" for active broadband single-ended probes, see "Copy to off-
set" on page 64. The offset value is also shown in the "Channel <n>" menu.
Com.Mode Offset
Sets the common mode offset to compensate for a common DC voltage applied to
both input terminals referenced to ground. You can measure the common mode input
voltage using the ProbeMeter.
ProbeMeter
®
Same as "ProbeMeter
®
" for active broadband single-ended probes, see "ProbeMeter
®
"
on page 64.
MicroButton
Same as for active single-ended probes, see "MicroButton" on page 64.
Meas. Mode
Sets the measurement mode of modular probes. You can switch between single-
ended, differential and common mode measurements without reconnecting or resolder-
ing the probe.
If you use the R&S RT-ZMA30 browser module, only DM measurements are possible
because this module has no ground connector.
The measurement modes are: