Remote control - status reporting system
R&S
®
RTA4000
822User Manual 1335.7898.02 ─ 08
Figure B-3: Overview of the STATus:QUEStionable register
Table B-4: Bits in the STATus:QUEStionable register
Bit No. Meaning
0 to 2 not used
3
COVerload
This bit is set if a questionable channel overload occurs (see Chapter B.3.4.1, "STATus:QUES-
tionable:COVerload register", on page 823).
4
TEMPerature
This bit is set if a questionable temperature occurs (see Chapter B.3.4.2, "STATus:QUEStiona-
ble:TEMPerature register", on page 823).