Applications
R&S
®
RTM3000
187User Manual 1335.9090.02 ─ 09
Min. Level
Defines the minimum absolute level for peak detection. Only peaks that exceed the
minimum level are detected, listed in the "Peak List", and marked in the spectrum dia-
gram.
Remote command:
SPECtrum:MARKer:SETup:MLEVel on page 551
Advanced Peak Setup
Enables a more precise peak detection.
For a peak to be detected, the Excursion, Max. Width and Distance have to be within
the defined limits.
Remote command:
SPECtrum:MARKer:SETup:MMODe on page 551
Max. Width
Defines the maximum width that a peak can have for it to be detected.
Remote command:
SPECtrum:MARKer:SETup:MWIDth on page 551
Excursion
Defines a minimum level difference between two subsequent peaks that has to be kept
for the peak to be detected. To avoid identifying noise peaks, enter a peak excursion
value that is higher than the noise levels.
Remote command:
SPECtrum:MARKer:SETup:EXCursion on page 551
Distance
Defines a distance between two subsequent peaks that has to be kept for the peak to
be detected.
Spectrum analysis and spectrogram (option R&S
RTM-K18)

Table of Contents

Related product manuals