Loading...
General Instrument Setup
R&S
®
Scope Rider RTH
285User Manual 1326.1578.02 ─ 12
Enable Touch
Switches the touch functionality of the screen on or off. Alternatively, you can press the
[SHIFT] key for 2 seconds.
In certain situations, the capacitive touchscreen can get stimulated without touching it.
Typical situations are:
Measuring in an environment with immissions considerably higher than specified
Measuring on power electronics with strong electric fields
Connecting the probe to power electronics DUTs with large voltages or fast rising
edges
Due to the unwanted stimulation, the R&S RTH detects three or more fingers on the
screen even though the display has not been touched. A message box informs about
the detection of more than two fingers. In this case, operate the instrument using the
keys and the wheel and disable the touchscreen.
Remote command:
DISPlay:MOUS on page 503
Selfalignment
See Chapter 13.3, "Selfalignment", on page 281.
Selftest
The self-test checks the hardware of the instrument. It is intended for service tasks.
USB/LAN
See Chapter 14.1, "LAN Connection", on page 292 and Chapter 14.2, "USB connec-
tion", on page 294.
Instrument Settings