Remote Control Commands
R&S
®
Scope Rider RTH
385User Manual 1326.1578.02 ─ 12
15.7.2.4 AC+DC Current and Voltage Measurements
METer<m>:SENSe:CURRent:AC:COUPling..................................................................... 385
METer<m>:SENSe:VOLTage:AC:COUPling...................................................................... 385
METer<m>:SENSe:CURRent:AC:COUPling <Coupling>
METer<m>:SENSe:VOLTage:AC:COUPling <Coupling>
Enables AC+DC measurements.
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Parameters:
<CoupVoltageMeas> DCLimit | ACLimit
DCLimit
Enables AC+DC measurement.
ACLimit
Enables AC measurement.
*RST: DCLimit
Example: Configure and perform AC+DC current measurement with range
100 A:
:METer:CONFigure:CURRent:AC 100
:METer:SENSe:FUNCtion 'CURR:AC'
:METer:CONFigure:VALue?
<-- "CURR:AC 100"
:METer:SENSe:CURRent:AC:COUPling DCL
:METer:READ?
<-- 0.035906488794
Configure and perform AC voltage measurement with range 10
V:
:METer:CONFigure:VAOLTage:AC 10
:METer:SENSe:FUNCtion 'VOLT:AC'
:METer:CONFigure:VALue?
<-- "VOLT:AC 10"
:METer:SENSe:CURRent:AC:COUPling ACL
:METer:READ?
<-- 0.030006488794
15.7.2.5 Temperature Measurements
METer:UNIT:TEMPerature...............................................................................................385
METer:SENSe:TEMPerature:TRANsductor:RTD:TYPE...................................................... 386
METer:UNIT:TEMPerature <Unit>
Sets the unit for temparature measurements.