Command Reference
R&S
®
ZNB/ZNBT
1283User Manual 1173.9163.02 ─ 54
Usage: Setting only
Manual operation: See "Start Cal Sweep" on page 557
SOURce<Ch>:POWer<PhyPt>:CORRection:NREadings <NoCalSweeps>
Sets a limit for the number of calibration sweeps in the source power calibration. The
command is valid for all channels, ports and external generators.
Equivalent command: SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:
AVERage[:COUNt].
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the limit
applies to all sources.
<PhyPt> Calibrated port number. This parameter is ignored; the limit
applies to all sources.
Parameters:
<NoCalSweeps> Number of readings
Range: 1 to 100
*RST: 2
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation: See "Flatness Cal – Max Iterations" on page 491
SOURce<Ch>:POWer<PhyPt>:CORRection:OSOurces[:STATe] <Boolean>
Switches off all other sources during the calibration sweep for channel <Ch>.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Calibrated port number. This parameter is ignored; the setting is
valid for all sources.
Parameters:
<Boolean> ON - other sources not necessarily switched off.
OFF - other sources switched off during the calibration sweep.
*RST: ON
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation: See "Switch Off Other Sources" on page 490
SOURce<Ch>:POWer<PhyPt>:CORRection:PPOWer <RecCalPower>
Defines the source power which the R&S ZNB/ZNBT uses to perform the first calibra-
tion sweep of the source power calibration ("Reference Receiver Cal Power"). The
power value is ignored if the R&S ZNB/ZNBT is set to use the port power result
(SOURce<Ch>:POWer<PhyPt>:CORRection:PSELect CPOW).
SCPI Command Reference