USER MANUAL - SIP POWER ION PUMP CONTROLLER
M.HIST.0109.23 Rev.1 Page 48 of 62
Bit 184:215 – Switch SW3 high threshold
This field contains switch SW3 high threshold in nA units.
Bit 216:247 – Keepalive interval
This field contains keepalive time interval in ms units. When keepalive interval is different from 0
ms the power supply expects to be polled within a period not greater than keepalive interval by
whom last sent to it the enable command, otherwise it stops and sets communication alarm latch.
Keepalive timer resets after any successfully processed command. Keepalive interval cannot be
lower than 1000 ms, the only exception is 0 ms which disables keepalive timer.
Bit 248:263 – Conversion rate
This field contains the conversion rate in A/Torr units useful to compute chamber pressure from
output current.
Bit 264:271 – Slave Modbus ID
When Modbus is available this field allows to change the slave Modbus ID (Allowed IDs: 1-247).
10.2. Payload for 0x41 – Set IP Address Command
Bit 0:31 – New IP address
This field contains the new Sip Power IP address.
Bit 32:63 – New IP network mask
This field contains the new Sip Power IP network mask.
10.3. Payload for 0x80 – Read All Answer Command
0 16bits 15 16 16bits 31 32 16bits 47 48 32bits 79 80 32bits 111
CARD_TYPE HW_CODE SW_VERSION SERIAL_NUMBER IOUT
112 16bits 127 128 16bits 143 144 16bits 159 160 16bits 175 176 16bits 191
VOUT VIN RESERVED TEMPERATURE ARCING_NUMBER
192 32bits 223 224 32bits 255 256 16bits 271 272 8bits 279 280 520bits 799
LIFE_TIME UPTIME STATUS SW_STATUS RESERVED
Bit 48:55 – Card type
This field is structured as follow:
• Bit 15:2 – Reserved bits
0 32bits 31 32 32bits 63
IP_ADDR IP_NETMASK