EasyManua.ls Logo

SAES SIP POWER - Page 49

SAES SIP POWER
62 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
USER MANUAL - SIP POWER ION PUMP CONTROLLER
M.HIST.0109.23 Rev.1 Page 49 of 62
These bits are reserved for future use.
Bit 1 – Ethernet flag
If this bit is set, it means that Sip Power features a Ethernet connection.
Bit 0 – Display flag
If this bit is set, it means that Sip Power features a display on the front panel.
Bit 16:31 – Hardware revision number
This field is structured as follow:
Bit 15:8 – Hardware revision major number
This is the hardware revision major number.
Bit 7:0 – Hardware revision minor number
This is the hardware revision minor number.
Bit 32:47 – Software version
Bit 15:8 – Software version major number
This is the software version major number.
Bit 7:0 – Software version minor number
This is the software version minor number.
Bit 48:79 – Serial number
This field shows the device serial number.
Bit 80:111 – Output current
This field shows output current in nA units (sampled in the same time slice of VIN and VOUT).
Bit 112:127 – Output voltage
This field shows output voltage in V units (sampled in the same time slice of VIN and IOUT).
Bit 128:143 – Input voltage
This field shows current input voltage in dV units (sampled in the same time slice of VOUT and
IOUT).
Bit 144:159 – Reserved
This field is reserved for future use.
Bit 160:175 – Internal temperature
This field contains the device internal temperature reading provided in Kelvin degrees.
Bit 176:191 – Arcing events number
This field contains the number of arcing events since last start (or restart) command.
Bit 192:223 – Total working hours
This field contains the device total working hours (only the time spent supplying current).
Bit 224:255 – Time elapsed since last start
This field contains the number of seconds elapsed from the last start (or restart).
Bit 256:271 – Power supply status
This field is structured as follow:
Bit 15:13 – Reserved bits