EasyManuals Logo

Sakura Cyto-Tek 2500 User Manual

Sakura Cyto-Tek 2500
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
Page 7
CAUTION: The instrument is intended to be operated
indoors. Do not use it outdoors.
COMPARISION OF ROTATION SPEED RELATIVE
TO CENTRIFUGAL FORCE
Rotation speed
(rpm)
Relative
Centrifugal
Force (g) Approx.
Acceleration
(m/s
2
)
Approx.
200
5
46
300
11
104
400
19
184
500
29
288
600
42
414
700
58
564
800
75
737
900
95
932
1000
117
1151
1100
142
1393
1200
169
1658
1300
198
1946
1400
230
2256
1500
264
2590
1600
301
2947
1700
339
3327
1800
380
3730
1900
424
4156
2000
470
4605
2100
518
5077
2200
568
5572
2300
621
6090
2400
676
6631
2500
734
7195
NOTES:
The distance from the center of the rotor to a
specimen chamber installed is 105 mm.
Relative centrifugal force (RCF) = 11.739 x 10
-5
x N
2
(g)
N = Revolutions per minute (rpm)

Other manuals for Sakura Cyto-Tek 2500