Page 31
5.3.1 Error Code of Monitored Errors
Error
Code
Description/Symptom Major Causes Action(s)
85 The instrument lid remains locked
Faulty solenoid or wiring
Contact Sakura Technical Support or distributor.
To remove specimen chambers, see “How to
unlock the lid mechanically during power outage”,
P2-15.
Poor connector contact
Controller board failure
Lid lock detection sensor failure
86 The instrument lid cannot be locked
Faulty solenoid or wiring
Contact Sakura Technical Support or distributor.
Poor connector contact
Controller board failure
Lid lock detection sensor failure
87
The instrument lid is opened during
operation.
Faulty lid sensor or improper wiring
Contact Sakura Technical Support or distributor.
Poor connector contact
Controller board failure
Lid lock detection sensor failure
88
The lock was released during
operation
Faulty solenoid or wiring
Contact Sakura Technical Support or distributor.
Poor connector contact
Controller board failure
Lid lock detection sensor failure
91
The instrument lid is open but the
solenoid is in the locked condition.
Faulty solenoid or wiring
Contact Sakura Technical Support or distributor.
Poor connector contact
Controller board failure
Lid lock detection sensor failure