3. OPERATING INSTRUCTIONS
Page 20
4. Press the START/STOP key (b) to start the operation
(Figure 3Z).
NOTE: The instrument lid is locked during operation.
CAUTION: The process time is selectable from 0
minute and 1 second to 99 minutes and 59 seconds.
However, a prolonged process time may make
specimens dry. Consider an appropriate process time
before setting the time.
CAUTION: After the start of operation, make sure that
the rotation speed reaches the set value and is
stable.
3.6 Discontinuing the Operation
To discontinue the operation, press the START/STOP key.
The instrument will start decelerating the motor and unlock
the instrument lid when the rotation speed reaches 50rpm
or less. An alarm will sound to the operator that the
instrument has stopped.
NOTE: While the lid is unlocked, the time left is still
displayed in the window. The operation, however, cannot
be restarted with the time left. When the lid is opened, the
instrument returns to the standby mode and resets the
displayed time to zero.
3.7 Emergency Stop
To stop the operation immediately, switch off the
instrument (Figure 3AA).
Figure 3AA
3.8 Removing the Guard Bowl
CAUTION: In order to minimize a risk of infection,
place the guard bowl inside a biological safety hood
when installing or removing the bowl lid.
1. When the operation has ended, a buzzer sounds to
inform the operator the run has completed. Open the
instrument lid and remove the guard bowl.
2. Place the guard bowl inside a safety cabinet. Pull up
the lock button to unlock the bowl lid and remove
(Figures 3BB and 3CC).
Figure 3BB
Figure 3CC
3. Remove the specimen chamber holders from the rotor.
4. Remove the glass slide from the chamber holder. Do
not to let the smeared area go out of alignment.
5. Drain the guard bowl of fluid waste. Disinfect or
sterilize the guard bowl and bowl lid.
Disinfect the 6mL or 12mL specimen chambers, chamber
caps, rubber gaskets, and 6/12mL chamber holders after
use. See Section 4.1.4 for cleaning instructions.
Dispose of the used 1mL specimen chambers, chamber
holders, chamber caps and filter paper, as appropriate