3 - 20 RS80A Service Manual
Date Format
Used to configure the date format. Select a format by using the combo
button. The date format that you specify will be applied to various date
fields in Patient Information.
Time
The current time is displayed.
Time Format
Select a time display format. Select a desired time format (12 Hour or 24
Hour) by pressing the combo button.
▐ Store Clip
Store Clip Method
Specify the method and range in which an image is acquired and saved.
You can select ECG Beat, Time or Manual. Note that ECG Beat can be
selected only when ECG is on.
- ECG Beat: Specify the heart beat as 1–8 beats.
- Time: Specify it as 1–600 seconds.
- Manual: Pressing the button on the control panel that has been
designated as Store Clip automatically starts saving the images;
pressing the same button again stops saving.
Cine Loop Period
- Prospective: When the Store Clip button is pressed during scanning,
the subsequent images are saved.
- Retrospective: When the Store Clip is pressed during scanning, the
previous images are saved.
NOTE
To configure the Store Clip button, go to Utility >
Setup > User Defined Key > User Key Setup.