c.Rotate the canister anti-clockwise, shift the canister’s three screws to the smaller ends of the
cone-shaped holes, then fix the screws with Inner Hexagon Wrench from the bottom of the
turntable.
C Change the nozzle
a. Switch the valve spanner(Fig. 1);
b. Remove the nozzle (Fig. 2), clean it for future use.
c. Clean the new nozzle. Install the O ring (8X1.8) (Fig. 3).
d. Switch the valve spanner, install the nozzle gently. (Fig. 4)
e. Pour out some paints after the reinstallation.
1 2
3 4
6.6 Calibration
Normally the machine has been precisely calibrated and ready for use before leaving the factory. Yet
in the following cases, new calibration is necessary:
When replacing either the Rough or Fine gauge or the location pin of the said gauge.
The dispensing result is not accurate.