-85-
Electrical Adjustments
Caution :
Other adjustment items are
not related to service and a
maintenance. Don't change
the data values of those
adjustment items. Otherwise
it may cause loss of product
safety.
Note : Group No. 930
■ If the data value appears "11" on the
screen, the lamp mode is selected.
■ If the data value appears "0" on the
screen, the lamp mode is not selected.
Set the data value to "10."
■ If the data value appears "99" on the
screen, it is period for cooling of the
lamp. Wait until it is changed to "0."
Then set the data value to "10."
Note : Fans control Group No. 140
Group Fans A: FN903, 904, 905, 916
Group Fans B: FN909, 910, 911, 912
Group Fans C: FN901, 906, 907,908
913, 914, 915
Group Fans D: FN917, 918, 919, 920
Not designated PC AV COMPONENT
1010 - 0 AV video signal adjustment (Green) ●
1010 - 1 AV video signal adjustment (Green) ●
1010 - 2 AV video signal adjustment (Red) ●
1010 - 3 AV video signal adjustment (Red) ●
1010 - 4 AV video signal adjustment (Blue) ●
1010 - 5 AV video signal adjustment (Blue) ●
0 - 0 Temperature monitor IC5801 Read only ●
0 - 1 Temperature monitor IC2541 Read only ●
1 - 0
Atmospheric Pressure monitor IC886 Read only ●
10 - 0 RS232C Baud rate (0:9600 / 1:19200) ●
10 - 1
RS232C RTS/CTS Flow Control (0:Disable / 1:Enable)
●
111 - 0 Lamp-1 Lamp lighting time Read only ●
111 - 1 Lamp-2 Lamp lighting time Read only ●
111 - 2 Lamp-3 Lamp lighting time Read only ●
111 - 3 Lamp-4 Lamp lighting time Read only ●
930 - 0 Lamp mode 4 Lamp Mode
(See Note.)
●
930 - 1 Lamp mode Lamp 1-4 of 2 Lamp Mode
(See Note.)
●
930 - 2 Lamp mode Lamp 2-3 of 2 Lamp Mode
(See Note.)
●
972 - 1 R_PANEL Ghost timer reset ●
972 - 3 G_PANEL Ghost timer reset ●
972 - 5 B_PANEL Ghost timer reset ●
998 - 0 Grid Test Pattern Display ●
Group-no. Adjustment Item Adjustment Mode
■ Test Points Location
Please refer to "Printed Wiring Board Drawings." for the locations of test point.